Calibration range and uncertainty finder

86 labs have scope examples you can screen by measurement, unit and range endpoints. The uncertainty filter compares fixed CMCs in the selected unit. For a formula, frequency band or field condition, open the scope to evaluate the entry at your test point.

Published scope examples

LabScope itemRangeCMCScope
Accredited Labs dba Huston-DePue Calibration Laboratory · Oklahoma City, OK DC Voltage – Generate (0 to 329.9999) mV 0.0054 % + 3 μV PDF p. 1
Accredited Labs dba Precision Calibration Systems · Morristown, TN Temperature Measuring (50 to 500) °C 1.5 % + 0.9 °C PDF p. 18
Accredited Labs dba Precision Calibration Systems · Morristown, TN Temperature – (-200 to 0) °C 0.020 °C PDF p. 19
Accredited Labs dba Precision Calibration Systems · Morristown, TN Temperature – (-90 to 40) °C 0.025 °C PDF p. 19
Advanced Calibration Technologies, dba Ad-Tek, Inc. · Boring, OR Digital Thermometers (-195 to 420) °C 0.029 °C PDF p. 7
All American Scales & Calibration · Paris, OH DC Voltage – Generate3 (0 to 330) mV 27 µV/V + 1 µV PDF p. 6
All American Scales & Calibration · Paris, OH Infrared Temperature – (35 to 100) °C 0.6 °C PDF p. 19
Applied Technical Services · Bartlett, TN Temperature – Measure3 (-80 to 420) C 0.056 C PDF p. 17
Applied Technical Services · Greer, SC Temperature – Measure3 (-80 to 420) °C 0.056 °C PDF p. 16
Applied Technical Services · Marietta, GA Temperature Measuring (-196 to 650) C 0.025 C PDF p. 45
Applied Technical Services · Marietta, GA Temperature – Measure3 (-196 to 650) C 0.025 C PDF p. 45
Applied Technical Services · Sacramento, CA DC Voltage – Generate3 (0 to 120) mV 8.7 µV/V + 0.88 µV PDF p. 4
Applied Technical Services · Sacramento, CA Temperature – Measure3 (-40 to 0) °C 1.2 °C PDF p. 16
Applied Technical Services · Sanford, FL Temperature – (-196 to 600) °C 0.056 °C PDF p. 16
Applied Technical Services · Tempe, AZ Temperature – Measure3 (-195 to 660) °C 0.056 °C PDF p. 17
Boise Calibration Service · Boise, ID Temperature – Measure (-100 to 0) °C 0.036 °C PDF p. 3
Calibration Solutions · Pflugerville, TX Temperature –Measure3 (-100 to 1000) °F 2.5 + 5.5 x 10-4T F PDF p. 4
Calibration Solutions · Pflugerville, TX Temperature – Measuring (32 to 700) °F (0.07 + 0.0014) °F PDF p. 4
Coast to Coast Calibrations · Blaine, MN DC Voltage – Generate3 (1 to 329) mV 0.0021 mV 0.000 PDF p. 7
Coast to Coast Calibrations · Blaine, MN Pressure Gages3 – (-12 to 100) psig 0.08 psi PDF p. 10
Coast to Coast Calibrations · Blaine, MN Hydraulic Pressure (200 to 2000) psig 1.6 psig PDF p. 11
Coast to Coast Calibrations · Blaine, MN Temperature (-20 to 40) °C 1.2 °C PDF p. 11
Coast to Coast Calibrations · Blaine, MN Thermometers & (-80 to -30) °C 0.025 °C PDF p. 11
Coast to Coast Calibrations · Blaine, MN Temperature Indicating (-30 to 100) °C 0.011 °C PDF p. 11
Coast to Coast Calibrations · Blaine, MN DC Voltage – Generate3 (1 to 329) mV 0.0021 mV PDF p. 18
CONNECTICUT CALIBRATION TECHNOLOGIES, INC., dba CONNECTICUT CALIBRATION LABS · Newtown, CT DC Voltage – Generate (0 to 330) mV 35 µV/V + 1 µV PDF p. 2
CONNECTICUT CALIBRATION TECHNOLOGIES, INC., dba CONNECTICUT CALIBRATION LABS · Newtown, CT Pressure Gages (1 to 300) psi 0.15 psi + 0.6R PDF p. 11
CONNECTICUT CALIBRATION TECHNOLOGIES, INC., dba CONNECTICUT CALIBRATION LABS · Newtown, CT Temperature – Measure (-35 to 350) °C 0.79 °C PDF p. 12
F. D. Hurka Metrology · Charlotte, NC DC Voltage – Generate3 (0 to 329.9999) mV 2.8 µV/V + 1.0 µV PDF p. 7
F. D. Hurka Metrology · Charlotte, NC Pressure – Measuring (0 to 15) psia 0.000 62 psia + 0.58R PDF p. 22
F. D. Hurka Metrology · Charlotte, NC Glass Thermometers, (0 to 232) °C 0.0091 °C + 0.58R PDF p. 25
F. D. Hurka Metrology · Charlotte, NC Temperature (-20 to 0.0) °C 0.0082 °C + 0.58R PDF p. 26
F. D. Hurka Metrology · Charlotte, NC Temperature – IR (-15 to 120) °C (0.43 + 0.0019t) + 0.58R PDF p. 26
FA Metrology, Formerly Known as FLORIDA METROLOGY, LLC · Port St. Lucie, FL Pressure Gages3, 4 (-14 to 0) psi 0.02 psi PDF p. 4
Fluke · Phoenix, AZ Temperature – Measure & (15 to 30) C 0.03 C PDF p. 4
Grand Rapids Metrology · Grand Rapids, MI DC Voltage – Generate3 (0 to 330) mV 20 µV/V + 1 µV PDF p. 3
Grand Rapids Metrology · Grand Rapids, MI Temperature – Measure3 (-200 to 0) °C 3.6 °C PDF p. 14
Grand Rapids Metrology · Grand Rapids, MI Temperature – (350 to 704) °C 3.3 °C PDF p. 15
Grand Rapids Metrology · Grand Rapids, MI Temperature – (-35 to 0) °C 0.13 °C PDF p. 15
Gulf Coast Calibration · La Porte, TX DC Voltage – Generate3 (0 to 329.9999) mV 15 µV/V + 1 µV PDF p. 2
Gulf Coast Calibration · La Porte, TX Pressure – Measuring (-15 to 1500) psi 0.30 psi PDF p. 7
Instrument Calibration Solutions · Aliquippa, PA Infrared Thermometers (35 to 500) °C 0.42 % + 0.33 °C PDF p. 19
JJ Calibrations · Portland, OR Temperature – Measure3 (-100 to 660) °C 0.005 °C PDF p. 13
JJ Calibrations · Portland, OR Temperature – (-30 to 250) °C 0.093 °C PDF p. 13
JJ Calibrations · Portland, OR Infrared Thermometers (-30 to 500) °C (0.092 + 0.000 32/°C) °C Infrared calibrator PDF p. 13
LAW Calibration · Saco, ME DC Voltage – Generate (0.00 to 120) mV 0.0009 % + 0.8 µV PDF p. 8
Midwest Metrology · Clayton, OH DC Voltage – Generate3 (0 to 220) mV 8 μV/V + 0.6 μV PDF p. 2
Midwest Metrology · Clayton, OH Hydraulic Pressure – (50 to 15 000) psi 0.025 % + 0.06 psi PDF p. 12
Midwest Metrology · Clayton, OH Infrared Temperature (35 to 100) °C 0.0028 °C/°C + 0.64 °C PDF p. 13
National Calibration · Albuquerque, NM Thermometers3 (-20 to 100) °C 0.059 °C PDF p. 13
National Calibration · Albuquerque, NM Temperature (0 to 50) °C 0.13 °C PDF p. 14
National Calibration · Scottsdale, AZ Temperature (0 to 70) °C 0.054 °C + 3 µ°C/°C PDF p. 25
National Calibration · Scottsdale, AZ Temperature – Measure3 (-80 to -20) °C 0.012 °C PDF p. 25
Northstar Calibration · Owatonna, MN DC Voltage – Generate3 (0 to 330) mV 0.000 36 % + 2.3 µV PDF p. 5
Northstar Calibration · Owatonna, MN Infrared Thermometers3 (85 to 750) °F (0.65 + 0.0016T) °F PDF p. 19
Northstar Calibration · Owatonna, MN Temperature – Measuring (-15 to 110) °C 0.91 °C PDF p. 19
Northstar Calibration · Owatonna, MN Temperature – Measure (-70 to -40) °C 0.91 °C PDF p. 19
Pacific Calibration Service · Campbell, CA Temperature – (-40 to 200) °C 0.11 °C PDF p. 2
Pacific Calibration Service · Campbell, CA Temperature – (-40 to 204) °C 0.11 °C PDF p. 2
PharmaMatrix, LLC dba Applied Calibration Services · Wilson, NC / Gauge Pressure (30 to 100) psi 0.01 % + 0.0014 psi PDF p. 8
PharmaMatrix, LLC dba Applied Calibration Services · Wilson, NC Temperature – Measure3 (-80 to 0) °C 0.046 °C PDF p. 9
PharmaMatrix, LLC dba Applied Calibration Services · Wilson, NC Temperature – (-80 to 0) °C 0.046 °C PDF p. 9
PharmaMatrix, LLC dba Applied Calibration Services · Wilson, NC Temperature – (-10 to 70) °C 0.065 °C PDF p. 9
Precision Measurements (ITS – An ATS Company) · West Chester, PA Pressure – Measuring (-15 to 25) psig 0.008 % of reading + PDF p. 17
Precision Measurements (ITS – An ATS Company) · West Chester, PA Pressure – Measuring 200 to 10 000) psig 0.000 83 % of reading PDF p. 18
Precision Measurements (ITS – An ATS Company) · West Chester, PA Thermometers (125 to 300) °C 0.036 °C PDF p. 18
Precision Repair And Calibrations · Blaine, MN Pressure Gages3 (10 to 10 000) psig 0.035 % PDF p. 20
Precision Repair And Calibrations · Blaine, MN Temperature – Measure (-180 to 0) °C 0.000 03 °C/°C + 0.026 °C PDF p. 22
Precision Repair And Calibrations · Blaine, MN Temperature – (-30 to 140) °C 0.000 072 °C/°C + 0.022 °C PDF p. 22
Quality Calibration Service · New Berlin, WI Pressure Gauges & (-13 to 300) psi 0.007 psi + 0.10 % PDF p. 15
Quality Calibration Service · New Berlin, WI Infrared Thermometers3 (-15 to 100) °C 0.74 °C PDF p. 18
Quality Testing Calibration (QTC) · Nicholasville, KY Pressure – Measure3 (5 to 300) psi 0.25 psi PDF p. 6
Quality Testing Calibration (QTC) · Nicholasville, KY Temperature – Measuring (-20 to 80) °C (0.05 + 0.000 25T) PDF p. 8
Rice Lake Weighing Systems · Concord, CA Temperature (20 to 400) C 0.15 C PDF p. 14
Rice Lake Weighing Systems · Concord, CA Temperature (-30 to 150) °C 0.15 °C PDF p. 14
Rice Lake Weighing Systems · Concord, CA PRT, Digital Thermometer (-80 to -30) °C 0.021 °C PDF p. 15
Riverview Calibration Services · Belpre, OH Temperature (0 to 200) °C 1.3 °C PDF p. 7
Riverview Calibration Services · Belpre, OH Temperature Probe (0 to 200) °C 1.3 °C PDF p. 7
Riverview Calibration Services · Belpre, OH Low Temperature (-40 to 40) °C 1.3 °C PDF p. 10
SIMCO · Allentown, PA DC Voltage – Generate3 (1.2 to 12) V 6.4 µV/V + 11 µV PDF p. 3
SIMCO · Allentown, PA Temperature – Measure3, (-50 to 150) °C 0.059 °C PDF p. 21
SIMCO · Bloomington, MN Temperature – Measuring (-20 to 140) °C 0.084 °C PDF p. 13
SIMCO · Bloomington, MN Temperature – Measure3 (-196 to 420) °C 0.068 °C PDF p. 13
SIMCO · Cordova, TN DC Voltage – Generate (0 to 330) mV 42 μV/V + 1 μV PDF p. 1
SIMCO · Greensboro, NC DC Voltage – Generate3 (0 to 330) mV 0.0021 % + 1 µV PDF p. 1
SIMCO · Greensboro, NC Pressure/Vacuum – (-15 to 30) psig 0.0093 psi PDF p. 13
SIMCO · Hampton, VA DC Voltage – Generate3 (0 to 220) mV 8.7 µV/V + 0.4 µV PDF p. 4
SIMCO · Hampton, VA Pressure – Measuring (0.7 to 50.8) psi 0.0016 % + 0.000 095 psi PDF p. 13
SIMCO · Hampton, VA Pressure Gauges – (100 to 10 000) psig 0.021 % PDF p. 13
SIMCO · Panorama City, CA Temperature – Measure3 (-196 to 0) °C 0.018 °C PDF p. 24
SIMCO · Panorama City, CA Temperature – (-75 to 175) °C 0.068 °C PDF p. 25
SIMCO · Redmond, WA DC Voltage – Generate3 (0 to 220) mV 9.0 V/V + 1.3 V PDF p. 3
SIMCO · Redmond, WA Pressure – Measuring (0 to 100) psia 0.026 % + 0.049 psia PDF p. 19
SIMCO · Redmond, WA Temperature – Measure (-196 to 420) °C 0.048 °C PDF p. 19
SIMCO · Redmond, WA Temperature – Measuring (-50 to 160) °C 0.13 °C PDF p. 19
SIMCO · Richardson, TX DC Voltage – Generate (0 to 0.22) V 9.4 µV/V + 0.4 µV PDF p. 16
SIMCO · Richardson, TX Temperature – Measure (-200 to -38) °C 0.025 °C PDF p. 19
SIMCO · San Diego, CA Pressure & Vacuum3 – (0.9 to 3.16) psia 0.000 38 psia + 0.58R PDF p. 12
SIMCO · Santa Clara, CA DC Voltage – Generate3 (0 to 220) mV 9 µV/V + 1.3 µV PDF p. 9
SIMCO · Santa Clara, CA Temperature – Measuring (-40 to 140) °C 0.027 °C PDF p. 15
SIMCO · Santa Clara, CA Temperature – Measure3 (-40 to 140) °C 0.028 °C PDF p. 16
SIMCO · Springboro, OH DC Voltage – Generate3 (0 to 330) mV 41 µV/V + 1 µV PDF p. 3
SIMCO · Springboro, OH Temperature – Measure3 (-196 to 420) °C 0.004 % + 0.037 °C PDF p. 13
SIMCO · Springboro, OH Infrared Thermometers3 (25 to 100) °C 0.68 °C PDF p. 13
SIMCO · Sterling, VA DC Voltage – Generate3 (0 to 120) mV 9.7 V/V + 0.8 V PDF p. 2
SIMCO · Sterling, VA Temperature – (-196 to 0) °C 0.090 °C PDF p. 24
Technology & Calibration, Inc. dba Tech Cal · Houston, TX Gauge Pressure – Hydraulic (5 to 1000) PSI 0.017 % of reading PDF p. 1
Tektronix · Irving, TX Infrared Temperature – (-30 to -20) °C 1.9 °C PDF p. 55
Total Calibration Solutions · Mentor, OH Pressure FO 100 to 150 psig 0.002 7 psig PDF p. 19
Transcat · Andover, MA DC Voltage – Source 1 (2.2 to 11) V 4 µV/V + 2.5 µV PDF p. 10
Transcat · Andover, MA Absolute Pressure Devices (50 to 150) psia 0.014 % of reading PDF p. 23
Transcat · Andover, MA Gage Pressure Devices (-15 to 25) psig 0.001 7 psi PDF p. 23
Transcat · Andover, MA Gage Pressure Devices (50 to 150) psig 0.012 % of reading + 0.012 psi PDF p. 23
Transcat · Andover, MA Temperature – Measure 1 (660 to 1 200) °C 0.46 % of reading + 2.9 °C PDF p. 24
Transcat · Charlotte, NC Pressure Devices (30 to 1 000) psia 0.007 % of reading + 0.000 5 psi PDF p. 28
Transcat · Charlotte, NC Pneumatic Gauge Pressure (-14.2 to 30) psig 0.002 psi PDF p. 28
Transcat · Charlotte, NC Temperature Measuring (300 to 420)°C 0.003% of reading + 0.35 °C PDF p. 30
Transcat · Cherry Hill, NJ Pneumatic Absolute Pressure (30 to 300) psia 0.008 8 % of reading PDF p. 34
Transcat · Cherry Hill, NJ Pneumatic Gauge Pressure (30 to 300) psig 0.007 5 % of reading PDF p. 34
Transcat · Cherry Hill, NJ Temperature – Measure 1 (600 to 1 000) °C 0.47 °C PDF p. 35
Transcat · Chesterfield, MO DC Voltage – Source 1 (1.2 to 12) V 0.000 62 % of reading + 7.8 µV PDF p. 12
Transcat · Chesterfield, MO Hydraulic Gauge Pressure (150 to 1 500) psig 0.008 % of reading + 0.02 psi PDF p. 28
Transcat · Chesterfield, MO Temperature – Source 1 (-20 to 25) °C 0.018 °C PDF p. 33
Transcat · Chesterfield, MO Infrared Thermometer 1 (120 to 200) °C 0.98 °C PDF p. 33
Transcat · Cincinnati, OH DC Voltage – Source 1,3 (3.3 to 33) V (0.000 003 + 0.000 015X) V PDF p. 7
Transcat · Cincinnati, OH Temperature – Measure 1,3 (-200 to 670) °C (0.003 + 0.000 15T) °C PDF p. 18
Transcat · Cincinnati, OH Thermometers 3 (-20 to 600) °C (0.08 + 0.000 9T) °C PDF p. 18
Transcat · Cincinnati, OH Temperature Sources 3 (-15 to 110) °C (0.3 + 0.000 3T) °C PDF p. 18
Transcat · Fullerton, CA Pressure Devices 1 (17 to 41.3) psia 0.003 psi PDF p. 36
Transcat · Fullerton, CA Pressure Devices 1 (41.3 to 1 000) psig 0.008 % of reading PDF p. 36
Transcat · Fullerton, CA Pressure Devices 1 (1 500 to 15 000) psig 0.008 % of reading PDF p. 36
Transcat · Fullerton, CA Temperature – Measure (-195 to 660) °C 0.003 % of reading + 0.009 °C PDF p. 37
Transcat · Houston, TX Pressure Devices (50 to 1 000) psia 0.002 2 % of reading PDF p. 67
Transcat · Houston, TX Pneumatic Pressure Devices (0.2 to 100) psig 0.001 7 % of reading + 0.000 008 psi PDF p. 67
Transcat · Houston, TX Hydraulic Pressure Devices (200 to 20 000) psig 0.003 % of reading + 0.007 3 psi PDF p. 67
Transcat · Houston, TX Temperature Measuring (100 to 300) °C 0.039 °C PDF p. 69
Transcat · Houston, TX Temperature – Measure (660 to 1 000) °C 8.7 °C PDF p. 69
Transcat · Indianapolis, IN Infrared Thermometers 1 (100 to 120) °C 0.72 °C PDF p. 33
Transcat · Mentor, OH DC Voltage – Source (3.3 to 33) V 10 µV/V + 16 µV PDF p. 3
Transcat · Mentor, OH Pressure Transducers 1,3 (160 to 1 300) psig 0.007 % of reading + 0.02 psi PDF p. 19
Transcat · Mentor, OH Temperature – Source 1,3 (-40 to 0) °C 0.068 °C PDF p. 22
Transcat · Mentor, OH (Temperature Measuring (0 to 160) °C 0.02 % of reading + 0.06 °C PDF p. 22
Transcat · Norfolk, VA Temperature – Measure 1 (0 to 100) °C 0.061 °C PDF p. 37
Transcat · Norfolk, VA Temperature – Measure 1 (420 to 960) °C 0.091 °C PDF p. 37
Transcat · Norfolk, VA Temperature – Source 1 (-30 to 0) °C 0.069 °C PDF p. 37
Transcat · Palm Beach Gardens, FL Hydraulic Pressure Devices (200 to 1 600) psig 0.091 psi PDF p. 30
Transcat · Palm Beach Gardens, FL Temperature – Measure 1 (155 to 420) °C 0.033 °C PDF p. 31
Transcat · Paxinos, PA DC Voltage – Source 1 (1.2 to 12) V 0.000 62 % of reading + 7.8 µV PDF p. 15
Transcat · Paxinos, PA Pressure Devices 1 (72.5 to 145) psig 0.012 % of reading + 0.008 3 psi PDF p. 20
Transcat · Rochester, NY DC Voltage – Source 1 (2.2 to 11) V 0.000 35 % of reading + 2.5 μV PDF p. 8
Transcat · Rochester, NY DC Voltage – Source 1 (220 to 1 100) V 0.000 65 % of reading + 0.4 mV PDF p. 8
Transcat · Rochester, NY Hydraulic Absolute Pressure (500 to 1 600) psia 0.011 % of reading + 0.07 psi PDF p. 32
Transcat · Rochester, NY Pneumatic Gauge Pressure (0 to 25) psig 0.001 4% of reading + 0.001 3 psi PDF p. 32
Transcat · Rochester, NY Hydraulic Gauge Pressure (500 to 1 600) psig 0.1 psi PDF p. 33
Transcat · Rochester, NY Temperature – Source 1 (600 to 800) °C 1.2 °C PDF p. 34
Transcat · Rochester, NY (Temperature Measuring (800 to 1 000) °C 1.5 °C PDF p. 34
Transcat · Rochester, NY Temperature – Measure 1 (0 to 420) °C 0.03 °C PDF p. 34
Transcat · San Diego, CA Infrared Thermometers (200 to 350) °C 1.9 °C PDF p. 24
Transcat · Somerset, PA DC Voltage – Source (2.2 to 11) V 4 µV/V + 2.5 µV PDF p. 3
Transcat · Somerset, PA Temperature – Measure 1 (0 to 420) °C 0.007 5 % of reading + 0.027 °C PDF p. 26
Transcat · Tempe, AZ Gauge Pressure – Source (-15 to 25) psig 0.001 7 psi PDF p. 37
Transcat · Tempe, AZ Gauge Pressure – Source (3 000 to 10 000) psig 0.01 % of reading + 0.25 psi PDF p. 37
Transcat · Tempe, AZ Temperature – Measure 1 (0 to 420) °C 0.025 °C PDF p. 38
Transcat · Tempe, AZ Infrared Temperature (120 to 200) °C 0.97 °C PDF p. 38
Transcat · Tigard, OR Pressure – Source 1 (0 to 1 500) psig 0.42 psi PDF p. 21
Transcat · Tigard, OR Temperature – Measure 1 (0 to 420) °C 0.026 °C PDF p. 22
Transcat · West Carrollton, OH DC Voltage – Source 1 (220 to 1 100) V 7.6 µV/V + 0.4 mV PDF p. 10
Transcat · West Carrollton, OH Pressure – Absolute 1 (30 to 300) psia 0.008 9 % of reading PDF p. 20
Transcat · West Carrollton, OH Pressure – Pneumatic 1 (30 to 300) psi 0.007 5 % of reading PDF p. 20
Trescal · Alpharetta, GA DC Voltage – Generate3 (0 to 220) mV 0.49 µV + 5.7 nV/mV PDF p. 2
Trescal · Alpharetta, GA Infrared Thermometers – (35 to 500) °C 0.54 °C + 0.0042 x TChange PDF p. 43
Trescal · Alpharetta, GA Temperature – Measure, (-197 to 660) °C 0.013 °C + 0.000 036 x PDF p. 43
Trescal · Arlington Heights, IL DC Voltage – Generate3 (0 to 220) mV 0.68 µV + 7 µV/V PDF p. 2
Trescal · Arlington Heights, IL Temperature – Measure3 (-196 to 420) °C 0.043 °C + 0.000 059 x PDF p. 31
Trescal · Arlington Heights, IL Infrared Temperature – (-15 to -12) °C 0.51 °C PDF p. 31
Trescal · Bohemia, NY Pressure –Measuring (-1 to 0) psig 0.013 psi + 0.6 R PDF p. 29
Trescal · Bohemia, NY Temperature – (-25 to 140) °C 0.06 °C PDF p. 30
Trescal · Bohemia, NY Temperature – (-196 to 660) °C 0.011 °C + ΔT(Change PDF p. 30
Trescal · Bohemia, NY Infrared Temperature – (35 to 500) °C 0.54 °C + ΔT(Change from 35 PDF p. 30
Trescal · Boonton, NJ Infrared Temperature – (35 to 500) °C 0.54 °C + 0.0042 x Δ PDF p. 33
Trescal · Cincinnati, OH DC Voltage – Generate3 (0 to 330) mV 0.8 µV + 16 nV/mV PDF p. 7
Trescal · Cincinnati, OH Temperature – (-196 to 420) °C 0.022 °C PDF p. 21
Trescal · Cincinnati, OH Temperature – (-38 to 100) °C 0.062 °C PDF p. 21
Trescal · Cornelius, NC Temperature – Measure3 (-196 to 0) °C 0.0055 °C - 0.000 004 x Additel 286 reference PDF p. 37
Trescal · ELIZABETH CITY, NC Temperature Uniformity (-200 to -190) °C 2.5 °C PDF p. 39
Trescal · Howell, MI DC Voltage – Generate3 (0 to 220) mV 0.4 µV + 7.5 µV/V PDF p. 8
Trescal · Howell, MI DC Voltage – Generate3 (0 to 120) mV 0.8 µV + 9.3 µV/V PDF p. 8
Trescal · Howell, MI Infrared Temperature – (-15 to -12) °C 0.51 °C PDF p. 79
Trescal · Lansdale, PA DC Voltage – Generate (0 to 330) mV 18 µV/V + 1 µV PDF p. 2
Trescal · Lansdale, PA Infrared Temperature – (-15 to -12) °C 0.51 °C PDF p. 13
Trescal · Lansdale, PA Temperature – Measure3 (-196 to 660) °C 0.0025 °C + 0.000 021 PDF p. 13
Trescal · Lansdale, PA Temperature – (0 to 100) °C 3.7 mK PDF p. 14
Trescal · Melbourne, FL DC Voltage – Generate3 (0 to 220) mV 0.4 µV + 7.5 µV/V PDF p. 6
Trescal · Melbourne, FL DC Voltage – Generate3 (0 to 120) mV 0.8 µV + 9.3 µV/V PDF p. 6
Trescal · Melbourne, FL Infrared Temperature – (-15 to -12) °C 0.51 °C PDF p. 40
Trescal · Milwaukee, WI DC Voltage – Generate (0 to 220) mV 5.6 μV/V + 0.39 μV PDF p. 24
Trescal · Milwaukee, WI Thermometers3 (35 to 500) °C 0.47 °C + 0.0047 °C/°C PDF p. 60
Trescal · Milwaukee, WI Temperature – (-196 to 660) °C 0.017 °C + 0.000 027 x PDF p. 60
Trescal · Oakwood Village, OH DC Voltage – Generate3 (0 to 220) mV 0.68 µV + 7 µV/V PDF p. 6
Trescal · Providence Forge, VA DC Voltage – Generate3 (0 to 330) mV 0.8 µV + 16 nV/mV PDF p. 2
Trescal · Providence Forge, VA Temperature – Measure3 (-70 to 180) °C 0.23 °C PDF p. 13
Trescal · San Jose, CA DC Voltage – Generate3 (0 to 220) mV 0.4 µV + 7.5 µV/V PDF p. 8
Trescal · San Jose, CA Temperature – Measure3 (-200 to 420) C 0.074 C PDF p. 56
Trescal · San Jose, CA Temperature – Measuring (-55 to 420) C 0.14 C PDF p. 57
Trescal · San Jose, CA Temperature: (0 to 40) °C 0.71 °C PDF p. 57
Trescal · Sparks, MD Temperature – Measuring (-45 to 140) °C 0.028 °C + 0.000 009 x PDF p. 20
Trescal · Tampa, FL Temperature – Measure (-200 to 660) °C 0.018 °C + 0.000 043 x PDF p. 17
Trescal · Tampa, FL Temperature – (-30 to 140) °C 0.071 °C + 0.000 096 x PDF p. 17
Trescal · Tampa, FL Temperature – (-30 to 260) °C 1.2 °C + 0.0015 x TChange from PDF p. 17
Trescal · Winston-Salem, NC Temperature – (-197 to 660) °C 0.0037 °C + 0.000 019 x ADT286-ST, 5699 SPRT PDF p. 10
Trescal · Winston-Salem, NC Temperature – (-80 to 300) °C 0.01 °C + 0.000 046 x PDF p. 10
Universal Calibrations · Jacksonville, FL Temperature – (-196 to 660) °C 0.08 °C PDF p. 2
Universal Calibrations · Jacksonville, FL Temperature – (-20 to 400) °C 0.09 °C + 0.6R PDF p. 2
usCalibration · Irvine, CA Temperature – Measure3 (-200 to 0) °C 0.065 °C PDF p. 19
usCalibration · Irvine, CA Temperature – (-25 to 0) °C 0.083 °C PDF p. 19
usCalibration · Irvine, CA Infrared Thermometers3 (32 to 216) °C 0.25 % + 0.92 °C PDF p. 19

Lab shortlist

Source data

Send your instrument model, quantity, test points, uncertainty requirement and due date. Include the worksite ZIP code for onsite service.

Request calibration quotes

Write it once. Calibration Service Companies passes it to matching labs, and their replies come straight to your inbox.

Done. Calibration Service Companies is routing your request to the right labs.

Calibration Service Companies forwards your request and contact details to matching labs so they can respond. Privacy policy

Questions buyers ask

How does the range filter work?

A row matches when its published interval contains your minimum and maximum in the selected unit. Enter the same value twice to screen one test point.

What does CMC mean?

Calibration and measurement capability is the uncertainty the lab can achieve under the conditions specified in the scope. The instrument under test and its conditions affect the uncertainty reported for your job. Source

How do I compare CMC formulas?

Calculate each formula at the same test point with the same units and scope conditions. Use the fixed CMC filter for rows that publish a single absolute value.